Julabo HighTech Ultra-Low Refrigerated & Heated 22 Liter Circulators

Category Stirred / Circulating Baths
Manufacturer Julabo
Country of manufacture Germany
Overview

Superior models for the most demanding applications
JULABO Ultra-Low Refrigerated Circulators for heating and cooling are suitable for external temperature tasks. Powerful circulating pump systems and high heating and cooling capacities guarantee short heat-up and cool-down times. ACC ‘Active Cooling Control‘ provides active cooling control across the whole temperature range. An energy saving feature with only minor loss of heat is provided by the proportional cooling capacity control in the ‘FP’ refrigeration units. The inevitable dust accumulation that occurs in refrigeration systems is solved by means of an easily removed venting grill, which allows for easy cleaning. The units are equipped with handles or castors for easy transportation. All models have a drain tap on the front allowing the fluid to be drained easily. Additionally the instruments are equipped with an improved insulation to help avoid ice-formation, and have a visual liquid level display. For working temperatures -91°C to +200 °C. Typical applications: Temperature applications to jacketed reaction vessels, autoclaves, miniplant installations, kilo labs, freezing point determination, calibration at low temperatures, petroleum testing, etc.

FPW91-SL is the most powerful JULABO ultra-low refrigerated circulator with an extended temperature range and 2-stage cascade cooling – suitable for high-volume reactors (up to 80 liters)

Models are available in the following options:

  • 90 = Unit size is 23.2 x 29.9 x 45.7 inches
  • 91 = Unit size is 33.5 x 29.9 x 45.7 inches
  • Models are available with a bath opening, suitable for external and internal application, unless designated with an N.
  • N = Models designated for external applications only offer a small filling nozzle to fill the bath liquid. No items can be immersed into the bath tank. Advantage is less cooling power loss, as there is no bath opening. (these models are designed with an N).
  • FP = Models which offer proportional cooling control to ensure energy-saving performance and to reduce heat generation to the environment.
  • W = Models are available in air or water-cooled. Water models are designated with a W. The advantages of water cooling are low noise level and almost zero emission of heat. But of course chilled water must be available with the customer.
  • 150 = Models are also available in an extended working temperature range of 150°C

  Advantages

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behaviour
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Early warning system for low liquid level (DBGM 203 06 059.8)
  • Adjustable high temperature cut-out, visible via display

Characteristics

Display  

A perfect view
Ample, easy to read VFD Comfort display for simultaneous display of 3 values, warning functions, high temperature cut-off, pump stages (resolution 0.01 °C)
Additional plain text information
Comfortable LCD dialog display for interactive operation with plain text display
Pump stage and liquid level
Backlit indicator for selected pump stages and filling volume on Presto® PLUS, Magnum 91 & Forte HT

Operation 

Comfortable and detailed
Comfortable keypad with additional menu functions for pump stages, calibration, control parameters, programmer, warnings, etc.

Temperature Control

For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal, <±0.05 °C external
Full control
Temperature Control Features’ for individual optimization, access to all important control parameters, additional settings for band limit, limits, Co-Speedfactor etc.
Highest measuring accuracy
Absolute Temperature Calibration’ for manual compensation of a temperature difference, 3-point calibration

Refrigeration Technology

Energy saving cooling (excludes F95-SL and FW95-SL models)
Proportional cooling control for automatic adjustment of cooling power or temporary switch-off of compressor as needed to save up to 90 % energy in comparison to unregulated cooling machines
Consistent cooling capacity
Easily removable venting grid for quick and easy cleaning
100 % Cooling capacity
Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Condensation and ice protection
A heated cover plate prevents condensation or ice build-up in the bath

Technical Features

Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Connection compliant to standard
RS232/RS485 dual-interface for serial data transmission according to EIA-485 industry standard (2-wire bus technology), upgradable with Profibus DP
Optimal program control
For the execution of time and temperature dependant profiles, 6 temperature profiles with 60 steps max., with real time clock
Automatic control of operating time
Electronic countdown-timer function for timer-programmed unit shut-down, standby mode after programmed time expires

Warning & Safety Functions

Early warning system for low liquid level
Maximum safety for applications, optical and audible alarm, allows user to refill bath fluid before the unit shuts down
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
Enhanced protective functions 
Maximum safety, adjustable high temperature cut-off or dry-running protection, additional display of setpoints permits easy and precise adjustments
For flammable bath fluids
Classification III (FL) according to DIN 12876-1

Key Attributes
Type Stirred / Circulating Baths Accessories
Incubation Yes
Refrigeration Yes
Temp Range (°C) Ambient  -95 to 150
Capacity (L) 22
Circulation External Only
Brand Julabo
MaxFlow Rate (l/min) 22-26
Temp Stability (°C) ±0.05
Timer Yes
Alarms Yes
Digital or Mechanical Digital
Data connection Yes
Tank Material Stainless Steel
Sample Agitation No
Media Racks Available No
Warranty (years) 1
Weight (lb) 430, 472
Width (in) 23.2
Depth (in) 29.9
Height (in) 45.7
Manufacturer Specifications

(Air Cooled)

 

Model FP90-SL FP90-SL FP90-SL-150C F95-SL
Sku No. 9352790 9352790N 9352790N150 9352795N
Model series HighTech HighTech HighTech HighTech
Working temperature range (°C) -90 ... +100 -90 ... +100 -90 ... +150 -95 ... 0
Temperature stability (°C) ±0.05 ±0.05 ±0.05 ±0.05
Setting / display resolution 0.01 °C 0.01 °C 0.01 °C 0.01 °C
Integrated programmer 6x60 steps 6x60 steps 6x60 steps 6x60 steps
Temperature Display VFD , LCD VFD , LCD VFD , LCD VFD , LCD
Heating capacity (kW) 3 3 3 3
Cooling capacity (Medium Ethanol) °C 20,   0,   -20,   -40,   -60,   -80 20,   0,   -20,   -40,   -60 20,   0,   -20,   -40,   -60,   -80 0,   -20,   -40,   -60,   -80
Cooling capacity (Medium Ethanol) kW 1.8,   1.7,   1.6,   1.35,   0.75,   0.15 1.8,   1.7,   1.6,   0.75,   0.15 1.8,   1.7,   1.6,   1.35,   0.75,   0.15 1.7,   1.5,   1.3,   1.1,   0.36
Pump capacity flow rate (l/min) 22-26 22-26 22-26 22-26
Pump capacity flow pressure (psi) 5.8-10.2 5.8-10.2 5.8-10.2 5.8-10.2
Pump capacity flow suction (psi) 2.9-5.8 2.9-5.8 2.9-5.8 2.9-5.8
Bath opening / bath depth (W x L / D inch) 11 x 9 None /  Depth 2.7 / 7.9 None /  Depth 2.7 / 7.9 None /  Depth 2.7 / 9.1
Pump connections M16x1 M16x1 M16x1 M16x1
Barbed fittings diameter (inner dia. / mm)  8 / 12  8 / 12  8 / 12  8 / 12
Filling volume (liters) 22 22 22 22
Refrigerant R23, R404A R23, R404A R23, R404A R23, R404A
External Pt100 sensor connection integrated integrated integrated integrated
Digital interfaces RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus
Ambient temperature 5...40 °C 5...40 °C 5...40 °C 5...40 °C
Dimensions W x L x H (inch) 23.2 x 29.9 x 45.7 23.2 x 29.9 x 45.7 23.2 x 29.9 x 45.7 23.2 x 29.9 x 45.7
Weight (LBS) 430 430 430 472
Classification according to DIN12876-1 Classification III (FL) Classification III (FL) Classification III (FL) Classification III (FL)
Included with each unit 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female).
Cooling of compressor Air Air Air Air
Power requirement V / Hz / A 3x 230/60/29   3x 230/60/29   3x 230/60/29   3x 230/60/28  
Additional information N/A N/A N/A N/A

 

(Water Cooled)

 

Model FPW90-SL FPW90-SL FPW90-SL-150C FPW91-SL FPW91-SL FW95-SL
Sku No. 9352791 9352791N 9352791N150 9352793 9352793N 9352796N
Model series HighTech HighTech HighTech HighTech HighTech HighTech
Working temperature range (°C) -90 ... +100 -90 ... +100 -90 ... +150 -91 ... +100 -91 ... +100 -95 ... 0
Temperature stability (°C) ±0.05 ±0.05 ±0.05 ±0.2 ±0.2 ±0.05
Setting / display resolution 0.01 °C 0.01 °C 0.01 °C 0.01 °C 0.01 °C 0.01 °C
Integrated programmer 6x60 steps 6x60 steps 6x60 steps 6x60 steps 6x60 steps 6x60 steps
Temperature Display VFD , LCD VFD , LCD VFD , LCD VFD , LCD VFD , LCD VFD , LCD
Heating capacity (kW) 3 3 3 3 3 3
Cooling capacity (Medium Ethanol) °C 20,   0,   -20,   -40,   -60,   -80 20,   0,   -20,   -40,   -60 20,   0,   -20,   -40,   -60,   -80 20,   0,   -20,   -40,   -60,   -80 20,   0,   -20,   -40,   -60,   -80 0,   -20,   -40,   -60,   -80
Cooling capacity (Medium Ethanol) kW 1.8,   1.7,   1.6,   1.35,   0.75,   0.15 1.8,   1.7,   1.6,   0.75,   0.15 1.8,   1.7,   1.6,   1.35,   0.75,   0.15 4.5,   4.1,   3.7,   3.1,   2,   0.75 4.5,   4.1,   3.7,   3.1,   2,   0.75 1.7,   1.5,   1.3,   1.1,   0.36
Pump capacity flow rate (l/min) 22-26 22-26 22-26 22-26 22-26 22-26
Pump capacity flow pressure (psi) 5.8-10.2 5.8-10.2 5.8-10.2 5.8-10.2 5.8-10.2 5.8-10.2
Pump capacity flow suction (psi) 2.9-5.8 2.9-5.8 2.9-5.8 2.9-5.8 2.9-5.8 2.9-5.8
Bath opening / bath depth (W x L / D inch) 11 x 9 None /  Depth 2.7 / 7.9 None /  Depth 2.7 / 7.9 11 x 9 None /  Depth 2.7 / 7.9 None /  Depth 2.7 / 7.9
Pump connections M16x1 M16x1 M16x1 M16x1 M16x1 M16x1
Barbed fittings diameter (inner dia. / mm)  8 / 12  8 / 12  8 / 12  8 / 12  8 / 12  8 / 12
Filling volume (liters) 22 22 22 22 22 22
Refrigerant R23, R404A R23, R404A R23, R404A R23, R404A R23, R404A R23, R404A
External Pt100 sensor connection integrated integrated integrated integrated integrated integrated
Digital interfaces RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus RS232, RS485, Optional Profibus
Ambient temperature 5...40 °C 5...40 °C 5...40 °C 5...40 °C 5...40 °C 5...40 °C
Dimensions W x L x H (inch) 23.2 x 29.9 x 45.7 23.2 x 29.9 x 45.7 23.2 x 29.9 x 45.7 33.5 x 29.9 x 45.7 33.5 x 29.9 x 45.7 23.2 x 29.9 x 45.7
Weight (LBS) 472 472 472 653 653 472
Classification according to DIN12876-1 Classification III (FL) Classification III (FL) Classification III (FL) Classification III (FL) Classification III (FL) Classification III (FL)
Included with each unit 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia.
Cooling of compressor Water Water Water Water Water Water
Power requirement V / Hz / A 3x 230/60/26   3x 230/60/32   3x 230/60/28   3x 230/60/32   3x 230/60/32   3x 230/60/28  
Additional information N/A Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. Cooling water connection G 3/4" male with barbed fittings for tubing 1/2" inner dia. N/A N/A
FAQ

Do the Julabo HighTech Circulators have the capability to ramp up temperature over a set time period?

The HighTech circulators can be programmed to ramp. The best way to accomplish this is to use the free version of Julabo's EasyTEMP software. EasyTEMP provides a PC interface to create a temperature profile which is much easier than programming directly circulator. The HighTech RS232 port enables connection to a PC using a null modem cable.

SHOW MORE SHOW LESS

select from options below to see SKU

Green Leaf
All shipments originating from Laboratory Supply Network are carbon neutral!

Do you need assistance?

We have experts standing by to assist you with your purchase. Call now at +1 603-652-1395 or contact us by clicking the button below.

Customer reviews