Cooling | SKU | Price |
---|---|---|
Air-Cooled | HBR:1000.0041.01 |
$28,000
|
Water-Cooled | HBR:1000.0042.01 |
$28,250
|
Air or Water Cooled | HBR:1000.0043.01 |
$30,000
|
The Unistat range inspires with unique thermodynamic properties and a range of functions to meet the highest demands. Unistat technology guarantees precise, reproducible temperature control results, the shortest possible heating and cooling times and large temperature ranges without the need for fluid change. All Unistats are fitted with the innovative Pilot ONE controller with a 5.7“ touchscreen color display and easy menu navigation. All important parameters can be seen at a glance and temperature curves are displayed in graphic real time. Unistats are equipped with the most up-to-date pump technology for optimal heat transfer. The pump self regulates to ensure the best possible circulation and flow and delicate glass reactors are protected from breakage by the pumps soft start-up feature. Unistat technology impresses as a result of economic operation with reduced operating costs and efficient energy management. Because Unistats are hydraulically sealed, the service life of the temperature control fluid is dramatically increased by preventing oil loss through vapors and oxidation, further saving resources. Unistats are fitted with extensive safety equipment and are therefore suitable for unsupervised continuous operation. Relevant operating conditions are constantly monitored and the system is switched off if necessary or with the process safety feature the emergency cooling circuit is activated.
The new multi-touch controller Pilot ONE offers a brilliant 5.7" touchscreen with graphic functions and a comfortable navigation menu. All important operating parameters and temperature values are neatly displayed on the touchscreen. Thanks to the new favorites menu, the one-click operation and the integrated technical glossary the operation is very easy, just like on a smartphone. Integrated USB and Ethernet ports allows connection to a PC or network, e.g. for remote control or data transmission. The E-grade Explore turns a Unistat into a development tool for process and chemical engineering.
Features:
Incubation | Yes |
Refrigeration | Yes |
Temp Range (°C) | -45 to 250 |
Capacity (L) | 1.5 |
Circulation | External |
Brand | Huber |
MaxFlow Rate (l/min) | 55 lpm ; 0.9 bar |
Temp Stability (°C) | ±0.01 |
Timer | Graph |
Alarms | Yes |
Digital or Mechanical | Digital |
Data connection | Yes |
Tank Material | N/A |
Sample Agitation | No |
Media Racks Available | No |
Warranty (years) | 1 year (extendable upon registration with manufacturer) |
Model | Unistat Tango | Tango w | Tango wL |
Temperature range | -45...250 °C | -45...250 °C | -45...250 °C |
Temperature stability | ±0.01 °C | ±0.01 °C | ±0.01 °C |
Temperature control | TAC, self-optimizing (True Adaptive Control) | TAC, self-optimizing (True Adaptive Control) | TAC, self-optimizing (True Adaptive Control) |
Display resolution | 0.01 °C | 0.01 °C | 0.01 °C |
Alarm message | optical, acoustical | optical, acoustical | optical, acoustical |
Controller | Pilot ONE (E-grade "Professional") | Pilot ONE (E-grade "Professional") | Pilot ONE (E-grade "Professional") |
Safety class | III / FL | III / FL | III / FL |
Protection class | IP20 | IP20 | IP20 |
Dimensions (W x D x H) | 16.77 x 12.87 x 24.84 inches | 16.77 x 12.87 x 24.84 inches | 16.77 x 12.87 x 24.84 inches |
Weight | 64 kg | 141.12 pounds | 54 kg | 119.07 pounds | 56 kg | 123.48 pounds |
Heating capacity | 3 kW | 3 kW | 3 kW |
Cooling capacity | 250 200 100 0 -20 -40 °C | 250 200 100 0 -20 -40 °C | 250 200 100 0 -20 -40 °C |
Cooling capacity | 0.7 0.7 0.7 0.7 0.4 0.06 kW | 0.7 0.7 0.7 0.7 0.4 0.06 kW | 0.7 0.7 0.7 0.7 0.4 0.06 kW |
Cooling machine | air-cooled | water-cooled | air- and water-cooled |
Refrigerant (ASHRAE, GHS) | R1270 (A3, H220) (0,15 kg) | R1270 (A3, H220) (0,09 kg) | R1270 (A3, H220) (0,15 kg) |
Pressure pump max. | 55 lpm ; 0.9 bar | 14.53 gpm ; 13.05 psi | 55 lpm ; 0.9 bar | 14.53 gpm ; 13.05 psi | 55 lpm ; 0.9 bar | 14.53 gpm ; 13.05 psi |
Pump connection | M24x1,5 AG male | M24x1,5 AG male | M24x1,5 AG male |
Permissible viscosity | 50 mm²/s | 50 mm²/s | 50 mm²/s |
Pump speed adjustable | Yes | Yes | Yes |
Pt100 connection | Pt100 | Pt100 | Pt100 |
Analogue input | via optional Com.G@te | via optional Com.G@te | via optional Com.G@te |
Analogue output | via optional Com.G@te | via optional Com.G@te | via optional Com.G@te |
Digital interface | RS232, Ethernet, USB Device, USB Host | RS232, Ethernet, USB Device, USB Host | RS232, Ethernet, USB Device, USB Host |
Permissible ambient temp. | 5...40 °C | 5...40 °C | 5...40 °C |
Noise level | 62 dB(A) | N/A | 62 dB(A) |
Filling volume min. | 1.5 Liters | 0.4 gal | 1.5 Liters | 0.4 gal | 1.5 Liters | 0.4 gal |
Volume of expansion | 2.8 Liters | 0.74 gal | 2.8 Liters | 0.74 gal | 2.8 Liters | 0.74 gal |
Cooling water connection | N/A | G1/2 AG | G1/2 AG |
Cooling water consumption | N/A | 24 l/h | 30 l/h |
Cooling water differential pressure min. | N/A | 0,5 bar | 72.5 psi | 0,5 bar | 72.5 psi |
Cooling water pressure max. | N/A | 6 bar | 87 psi | 6 bar | 87 psi |
Power supply | 208V 2~ 60Hz | 208V 2~ 60Hz | 208V 2~ 60Hz |
Current max. | 18 A | 16,5 A | 17,5 A |
Fusing | 20 A | 2x20 A | 2x20 A |